Method Article
Herein we report a tailored HPLC purification protocol that yields high-purity amyloid beta 42 (Aβ42) and amyloid beta 40 (Aβ40) peptides, capable of oligomer formation. Amyloid beta is a highly aggregation prone, hydrophobic peptide implicated in Alzheimer's disease. The amyloidogenic nature of the peptide makes its purification a challenge.
Amyloidogenic peptides such as the Alzheimer's disease-implicated Amyloid beta (Aβ), can present a significant challenge when trying to obtain high purity material. Here we present a tailored HPLC purification protocol to produce high-purity amyloid beta 42 (Aβ42) and amyloid beta 40 (Aβ40) peptides. We have found that the combination of commercially available hydrophobic poly(styrene/divinylbenzene) stationary phase, polymer laboratory reverse phase - styrenedivinylbenzene (PLRP-S) under high pH conditions, enables the attainment of high purity (>95%) Aβ42 in a single chromatographic run. The purification is highly reproducible and can be amended to both semi-preparative and analytical conditions depending upon the amount of material wished to be purified. The protocol can also be applied to the Aβ40 peptide with identical success and without the need to alter the method.
알츠하이머 병은 신경 퇴행성 질환 효과 만 35 명 이상의 세계이다. 질병의 발병 및 발달에 크게 관여 1은 매우 응집하기 쉬운 소수성 펩타이드 베타 아밀로이드 (Aβ)이다. 이 Aβ가 그러나, 아밀로이드 베타 42 (Aβ42)는 단백질의 가장 독성 형태 42 아미노산 변이체 인 것으로 생각하고, 길이 36 내지 43 개의 아미노산의 범위이다. (3)이 용이 특히 신경 독성 엔티티 것으로 여겨진다 확산 성, 올리고머 성 종을 형성하는 Aβ42의 능력에 대부분 기인한다. 도 4는 Aβ 펩티드에 대한 이해를 증진하기 위하여, 일상적으로 고순도 재료를 얻는 것이 필수적이다. 미량 불순물의 존재는 크게 펩티드 응집 성향의 특성을 변화하는 것으로 나타났다. (5)
티raditionally 같은 소수성 Aβ 펩티드의 고속 액체 크로마토 그래피 (HPLC) 분리 C 사 또는 C 8 실리카 계 정지상과 이동상 산성의 조합을 사용하여 수행되었다. (6) 그러나, 이러한 조건은 펩티드의 정제에 도전을 제공 할 수있다. Aβ 펩티드의 낮은 등전점은 PI (약 5.5) (7)는 산성 조건 하에서, 펩티드의 응집이 증가되고, 그 결과 종종 분리하기 어려운 넓은 비 - 분해 HPLC 피크는 (도 2A)을 생산하는 것을 의미한다. 또한, 이러한 넓은 피크들은 펩티드의 응집 프로필 영향 및 일반적 극적 제조 펩타이드의 양에 영향을 미칠 수있는 정제의 후속 발사가 필요할 수 불순물을 함유한다.
폴리 (스티렌 / 디 비닐 벤젠) 정지상 PLRP-S는을 정제의 대안적인 수단을 나타낸다잉 소수성 펩타이드. 정지상 다른 단백질과 메신저 리보 핵산 (mRNA의) 다수의 정제에 사용되어왔다. 8, 9 PLRP-S 정지상은 역상 분리 추가적인 알킬 리간드를 필요로하지 않고,보다 중요한 펩티드 deaggregation 리드 높은 pH에서 화학적으로 안정하다. 7 여기서, 우리는 고순도 아밀로이드 베타 42 (Aβ42)과 아밀로이드 베타 40 (Aβ40) 펩타이드를 산출 맞춤형 HPLC 정화 프로토콜을보고합니다.
1. Preparative HPLC Purification of the Aβ40 or Aβ42 Peptide
Figure 1: Experimental setup of the HPLC instrument used for purification of the amyloid beta peptides. (A) The quaternary HPLC pump fitted with a degasser and variable wavelength detector set to 214 nm and 280 nm; (B) HPLC columns used for purification of the amyloid beta peptides, from left to right, 25 x 300 mm2 preparative column, 7.5 x 300 mm2 semi preparative column and 4.6 x 250 mm2 analytical column; (C) Manual injector with 20 µL stainless steel injection loop used for analytical HPLC; (D) Manual injector with 10 mL stainless steel injection loop used for preparative and semi preparative purification. Please click here to view a larger version of this figure.
Time / min | % of Buffer Aa | % of Buffer Bb | Flow Rated / mL min-1 |
0 | 80 | 20 | 6 |
45 | 75.5 | 24.5 | 6 |
45.01 | 80 | 20 | 6 |
52.01 | 80 | 20 | 6 |
52.02 | 73 | 27 | 6 |
85 | 73 | 27 | 6 |
92 | 5 | 95c | 6 |
Table 1: Timetable for the purification of the Aβ42 and Aβ40 peptides using the 25 × 300 mm polymer column. aBuffer A - H2O with 20 mM NH4OH; b80% MeCN / 20% H2O with 20 mM NH4OH; cRan for 15 min to wash the column prior to the next injection of sample; dIn order to run a flow rate of 6 mL/min on the HPLC instrumentation, the pressure limit needs to be reduced to 200 bar.
2. Analytical HPLC Analysis of the Purified Aβ Protein
Time / min | % of Buffer Aa | % of Buffer Bb | Flow Rate / mL min-1 |
0 | 95 | 5 | 1 |
30 | 50 | 50 | 1 |
Table 2: Timetable for the HPLC purity analysis of the Aβ peptide. aBuffer A - H2O with 20 mM NH4OH; b80% MeCN / 20% H2O with 20 mM NH4OH.
Figure 2: Representative HPLC traces of Aβ42. (A) Traditional C4 silica purification, conditions: Buffer A: H2O with 0.1% trifluoroacetic acid (TFA), buffer B: MeCN (acetonitrile) with 0.1% TFA, gradient: 20 to 27% buffer B over 40 min followed by isocratic 27% buffer B; (B) Preparative purification using the 25 x 300 mm2 polymer column, conditions: Buffer A: H2O with 20 mM NH4OH, buffer B: 80% MeCN / 20% H2O with 20 mM NH4OH, gradient: 20 to 27% buffer B over 70 min followed by isocratic 27% buffer B; (C) Optimized preparative purification using the 25 x 300 mm2 polymer column, conditions are described in Table 1 located in sub-section 1.3 of the protocol description text; (D) Analytical HPLC using the 4.6 x 250 mm2 polymer column, conditions: Buffer A: H2O with 20 mM NH4OH, buffer B: 80% MeCN / 20% H2O with 20 mM NH4OH, gradient-5 to 50% buffer B over 30 min. For parts A, B and C the peak corresponding to Aβ42 is marked by an asterisk. Collection of the marked Aβ42 peak in part C reveals a purity of >95% as shown in part D. Mass spectrometry was used to determine the identity of the Aβ42 peak. Please click here to view a larger version of this figure.
The purification of the Aβ42 peptide using a combination of the PLRP-S stationary phase and a high pH mobile phase results in the formation of a sharp, resolved peak for the Aβ peptide at a retention time between 72 and 74 min (Figure 2C). Confirmation of the identity of the peak is done through direct injection mass spectrometry of the collected eluent. The eluent can be stored at -20 °C in solution for up to 12 h. Longer periods of storage may result in oxidation of the protein. To isolate the purified peptide, the eluent is flash frozen in liquid nitrogen and lyophilized. Analytical HPLC of the purified peptide reveals a purity of >95% (Figure 2D). The same procedure can be used to purify the Aβ40 peptide (Figure 3A) without modification of the method. Analytical HPLC indicates a purity of Aβ40 to be greater than 95% (Figure 3B).
Figure 3: Representative HPLC traces of Aβ40. (A) Optimized preparative purification using the 25 x 300 mm2 polymer column, conditions are described in Table 1 located in sub-section 1.3 of the protocol description text. The peak corresponding to Aβ40 is marked with an asterisk. (B) Analytical HPLC using the 4.6 x 250 mm2 polymer column, conditions: Buffer A: H2O with 20 mM NH4OH, buffer B: 80% MeCN / 20% H2O with 20 mM NH4OH, gradient: 5 to 50% buffer B over 30 min. Collection of the marked Aβ40 peak in part A reveals a purity of >95% as shown in part B. Mass spectrometry was used to determine the identity of the Aβ40 peak. Please click here to view a larger version of this figure.
In order to confirm that the purified peptide can form oligomeric mixtures, we performed photo-induced crosslinking of unmodified proteins (PICUP) as previously described14,15 and were able to robustly observe oligomer formation.10
The HPLC purification of the Aβ peptide is highly dependent upon the choice of both the stationary phase employed in the purification and the mobile phase chosen to elute the peptide. The low isoelectric point of the peptide and high propensity for aggregation render traditional chromatographic conditions for the separation of hydrophobic proteins (C4 or C8 stationary phase coupled with an acidic mobile eluent) challenging, with the Aβ peptide eluting as a prolonged broad, non-resolved peak (Figure 2A).
To circumvent this issue, the PLRP-S stationary phase, chemically stable at high pH was found to be effective for the purification of the Aβ peptide (Figure 2B and 2C). The use of a high pH mobile phase minimized the degree of aggregation, and when optimized, gave rise to a sharp, well-resolved peak. In the optimized protocol, the percentage of buffer B was rapidly switched from 20% to 27% at the 52-minute time point and as a result, gave rise to a well-defined Aβ peak (Figure 2C). This optimized protocol relies on all of the hydrophobic impurities being eluted from the column during the initial 20 to 24.5% increase in buffer B concentration. If impurities are found which are of a similar hydrophobicity as the Aβ peptide, then further optimization of the HPLC protocol may be warranted. The optimized method was highly reproducible amongst individual batches of synthesized Aβ and was capable of purifying both 40 and 42 variants of the peptide without any changes to the optimized procedure (Figure 3). For both peptides, the purity as determined by analytical HPLC was found to be greater than 95%.
Given the reports of trace impurities altering the aggregation propensity of the Aβ peptide, it is advisable that an orthogonal biophysical characterization be employed to confirm that the purified peptide is able to undergo oligomerization. Using previously reported protocols we chose to perform PICUP.14,15 PICUP analysis of the purified proteins demonstrates the characteristic monomer through hexamer population distribution for the Aβ40 peptide and the monomer through heptamer distribution associated with the Aβ42 peptide.10 These results confirm that purification of the Aβ peptides using the PLRP-S stationary phase and a high pH mobile eluent results in Aβ proteins that are capable of aggregation.
The purification procedure reported is designed to be able to purify up to 3 mg of the Aβ peptide in a single chromatographic run. For this procedure, it is critical to set the flow rate of the HPLC instrument at 6 mL/min. If lower flow rates of the HPLC pump are used, the HPLC run time should be extended to accommodate this. Conversely, the use of higher flow rates may warrant a reduction in the HPLC run time. However, reduction of the run time can result in the co-elution of the Aβ peak with other peaks and therefore lower the purity of the Aβ peptide collected. Furthermore, the HPLC instrumentation and size of the chromatographic column used for purification can greatly affect the amount of material that can be purified in a single run. If no peak is produced at the expected eluent time of the Aβ peptide, and a large peak is produced at the eluent time corresponding to the solvent front, then too much material was initially loaded onto the column. Reducing the amount of material injected onto the HPLC column for each run will circumvent this problem. Once the Aβ peptide has been purified, it is highly recommended that the solution be lyophilized as quickly as possible. Prolonged storage of the peptide in solution can lead to oxidation of Aβ and therefore formation of impurities. The purity of the Aβ peptide should always be determined by analytical HPLC. Trace amounts of impurities can alter the aggregation propensity of the peptide dramatically. If the analytical HPLC trace shows the presence of impurities, then the sample should be re-subjected to the HPLC purification protocol and its purity re-determined by analytical HPLC.
It is hoped that this tailored procedure for the purification of the Aβ42 and Aβ40 peptides will be of benefit for the scientific community and allow users to obtain high purity Aβ capable of oligomer formation. It would be expected that this procedure could be adapted to other amyloidogenic peptides that are difficult to isolate and purify.
The authors have nothing to disclose.
The authors would like to thank Agilent for their technical assistance. Kate Markham and Rafael Palomino are credited for their initial help in the synthesis and purification of the Aβ peptide and Dr Hsiau-Wei Lee is thanked for his help in preparing Figure 1 of the manuscript.
Name | Company | Catalog Number | Comments |
Agilent 1260 Infinity II quarternary pump | Agilent | G7111B | http://www.agilent.com/en-us/products/liquid-chromatography/lc-pumps-vacuum-degassers/1260-infinity-ii-quaternary-pump |
Agilent 1260 Infinity II Dual variable wavelength detector | Agilent | G7114A | http://www.agilent.com/en-us/products/liquid-chromatography/lc-detectors/1260-infinity-ii-variable-wavelength-detector |
Agilent 1260 Infinity II Manual Injector fitted with 10 mL stainless steel sample loop | Agilent | 0101-1232 | http://www.agilent.com/en-us/products/liquid-chromatography/lc-injection-systems/1260-infinity-ii-manual-injector |
Agilent 1260 Infinity II Manual Injector fitted with 20 µL stainless steel sample loop | Agilent | G1328C | http://www.agilent.com/en-us/products/liquid-chromatography/lc-injection-systems/1260-infinity-ii-manual-injector |
Ring Stand Mounting Bracket | Agilent | 1400-3166 | |
Agilent PLRP-S 300 Å 5 µm 4.6 x 250 mm (Analytical) | Agilent | PL1512-5501 | http://www.agilent.com/en-us/products/liquid-chromatography/lc-columns/biomolecule-separations/plrp-s-for-biomolecules#features |
Aβ42 or Aβ40 peptide | Synthesized in-house using a CEM liberty automated peptide synthesizer. | ||
Ammonium Hydroxide (NH4OH, 28% solution) | Fisher Scientific | A669-500 | |
Acetonitrile | Fisher Scientific | A998-4 | |
HPLC grade water | Fisher Scientific | W5-4 | |
Falcon 50 mL conical centrifuge tube | Fisher Scientific | 14-954-49A | |
Supelco PEEK Fitting One-piece fingertight, pkg of 5 ea | Sigma-Aldrich | Z227250 | |
Normject 5 cc sterile syringe | Fisher Scientific | 1481729 | |
16 Gauge SS Needle | Rheodyne | 3725-086 |