Method Article
Microfluidic double emulsions generation typically involves devices with patterned wettability or custom-fabricated glass components. Here we describe the fabrication and testing of an all polydimethylsiloxane (PDMS) double emulsion generator that does not require surface treatment or complicated fabrication processes, and is capable of producing double emulsions down to 14 µm.
Double emulsions are useful in a number of biological and industrial applications in which it is important to have an aqueous carrier fluid. This paper presents a polydimethylsiloxane (PDMS) microfluidic device capable of generating water/oil/water double emulsions using a coaxial flow focusing geometry that can be fabricated entirely using soft lithography. Similar to emulsion devices using glass capillaries, double emulsions can be formed in channels with uniform wettability and with dimensions much smaller than the channel sizes. Three dimensional flow focusing geometry is achieved by casting a pair of PDMS slabs using two layer soft lithography, then mating the slabs together in a clamshell configuration. Complementary locking features molded into the PDMS slabs enable the accurate registration of features on each of the slab surfaces. Device testing demonstrates formation of double emulsions from 14 µm to 50 µm in diameter while using large channels that are robust against fouling and clogging.
Double emulsions consist of droplets separated from a carrier phase by an intermediate, immiscible fluid layer, and are of particular interest due to their potential uses in industrial, pharmaceutical, and biological applications1. In some cases, the ability to encapsulate high value compounds in a double emulsion's core enables material to be protected and released in a controlled manner. For example, drugs may be encapsulated under solubility conditions not appropriate for the external carrier fluid2. Additionally, the intermediate oil layer can be used as a capsule template for the encapsulation and delivery of drugs, cosmetics, and nutrients3. In biology, double emulsions are also useful in high throughput screening because they allow a massive number of sub-nanoliter experiments to be carried out, then detected and sorted using a fluorescence-activated cell sorting (FACS) instrument4,5.
The design of double emulsions with desired performance characteristics requires the precise control of double emulsion size, composition, and uniformity. Although bulk emulsification processes, such as membrane emulsification, are used in industry, the resulting emulsions are highly polydisperse, exhibiting a wide variety of functional properties1. The field of droplet microfluidics is naturally suited the generation of monodisperse emulsions with carefully controlled composition6. Microfluidic double emulsion generation has been achieved with two main strategies, sequential drop making and glass capillary flow focusing. Double emulsions can be generated in planar PDMS devices using a two-step drop making process. First, aqueous-in-oil emulsions are created using a water-in-oil drop-making region of a device with hydrophobic channel walls. Next, the emulsion can be flowed or reinjected into a drop-making region with hydrophilic walls suited for oil in water drop-making4. However, hydrophilic surface treatment of PMDS requires an additional fabrication step and is often temporary7. The most controllable and repeatable method to form double emulsions is by co-axial flow focusing, a technique pioneered using glass capillary microfluidics, whereby a concentric jet containing the three phases is sheared through a small orifice to produce monodisperse droplets8. This technique allows for the production of droplets much smaller than the channel dimensions, with the precise size and composition of the double emulsion being a function of the flow rates of each phase. The large difference between droplet and channel size and the protective outer sheath flow prevents droplets from contacting the channel walls, rendering surface treatment unnecessary. However, such glass devices require custom fabrication of tapered capillary tips, along with careful assembly and sealing. Previous investigators have used 3D soft lithography to generate double emulsions using flow focusing physics, but these devices produced emulsions with diameters > 150 µm 9,10, roughly an order of magnitude larger than objects typically sorted with FACS. An attractive alternative would include the robust functionality and small droplet generation of glass capillary coaxial flow focusing with the ease of manufacture of PDMS soft lithography.
In this paper, we describe a double emulsion generator that uses co-axial flow focusing to produce ≤ 50 µm emulsions and is constructed entirely using 3D soft lithography11. Our device uses a clamshell approach to fabricate devices that includes a small shearing channel (Figure 1) to approximate the emulsion formation processes in a pulled glass capillary nozzle. More importantly, these devices require no specific surface treatment, and the all polymer construction provides easy and repeatable fabrication scalable to a large number of duplicate devices. Here, we outline the design, fabrication, and testing of the double emulsion generator. Double emulsion generation is shown to be robust and repeatable down to droplet diameters of 14 µm. The coupling of functionality with ease of fabrication makes this device an appealing option for development of new double emulsion applications.
1. SU8 Master Fabrication
2. PDMS Device Fabrication
3. Preparation of Reagents
4. System Preparation
5. Emulsion Generation
The double emulsion generator consists of a co-axial flow focusing device created using 3D PDMS fabrication (Figure 1A). The geometry enables that formation of a three-phase co-axial jet to be sheared into a square, 50 µm x 50 µm orifice, allowing the formation of water / oil / water double emulsions (Figure 1B, Figure 1C). The inner aqueous phase and the middle oil phase are brought together at a junction with channel dimensions of 10 µm x 50 µm (Figure 1D, point "1"). Due to the hydrophobicity of PDMS, the fluorinated oil hugs the channel walls and the inner phase stays in the center of the channel as the fluids travel in a continuous jet, until a sudden expansion channel expansion is reached (Figure 1D, point "2"). At this location, the inner two phases are injected into the center of a 320 µm tall junction that allows the relatively concentric introduction of the aqueous carrier phase. The three phases are forced into a 50 µm x 50 µm orifice (Figure 1D, point "3"), whereby the high flow rate of the carrier phase shears the inner two phases into a long, thin tendril that decomposes to from uniform droplets (Figure 1E).
The 3D PDMS fabrication requires the coupling of two unique PDMS molds in a clamshell configuration after molding on two layer lithographic masters. A 50 µm tall layer is used to form the inner and middle fluid handling channels, along with the shearing orifice on Master 1 (Figure 2A), along with a complimentary protruding and recessed frames on opposing master. An additional 135 µm tall layer is used to create the carrier fluid and the exit channels (Figure 2B). Assembly of the double emulsion generator utilizes the recessed and protruding frames (Figure 3A) for geometric alignment after plasma treatment (Figure 3B).
The double emulsion device was tested at a variety of flow conditions to demonstrate the formation of varied size, monodisperse double emulsions. For these experiments, the inner and middle phase flow rates were held constant and the carrier phase flow rate was modified to affect the shearing force during droplet generation. Experimental conditions are parameterized by the ratio of the carrier phase flow (Qc) to the sum of the inner two phases flows (Qsum). Images of droplet generation for experiments performed at Qc/Qsum from 3 to 57 are shown in Figure 4. An elongated region containing the inner two phases is observed to protrude into the 50 µm x 50 µm orifice and breaks into droplets that are convected downstream. Increasing the flow of the carrier phase (increasing Qc/Qsum) leads to the inner phases being sheared into progressively thinner regions that produce smaller droplets. Double emulsions produced by the device at different flow rates show an average diameter coefficient of variation of 5.2%. Histograms of droplet diameters for select values of Qc/Qsum also show the relative uniformity in the size of the droplets generated (Figure 5). The device demonstrates an ability to form double emulsions significantly smaller the orifice width, and shows a clear decreasing trend with increased Qc/Qsum (Figure 6). At highest carrier phase flow tested, 14 µm double emulsions were formed using the 50 µm x 50 µm orifice.
Figure 1. Geometry of the double emulsion generator. (A) 3D model of the fabricated device. (B) Vertical cross section of the central channel showing introduction of the inner (gray), middle (red), and carrier (blue) phases. (C) Cross-section showing the jet containing the inner two phases entering the square orifice. (D) Top view of emulsion generation in the device. At junction (1) the injection of the hydrophobic middle phase is aided by the hydrophilic PDMS, which causes it to coat the channel walls. At junction (2) the channel expands and a jet of the inner two phases is sheared into the orifice (3) by the higher rate of flow of the continuous fluid to a point where the physics cause droplet formation. (E) A microscope image of double emulsion generation in the device. Please click here to view a larger version of this figure.
Figure 2. Lithographic production of the masters. (A) The mask used for preparation of 50 µm features. Master 1 is used to mold the fluidic inlets, the inner / middle phase junction, the emulsion generation orifice, and a recessed trough for alignment. Master 2 contains a raised ridge used for alignment. (B) The mask used for preparation of 135 µm features. The masters are mirror images that contain the carrier fluid routing channels and the exit channel. Please click here to view a larger version of this figure.
Figure 3. Assembly of the PDMS device. (A) Fabricated PDMS slabs contain complimentary recessed and protruding frames. (B) Assembled, the frames interlock to provide optimum alignment of features. Please click here to view a larger version of this figure.
Figure 4. Images of generated double emulsions at different flow rates. The flow rate of the outer phase is changed to alter Qc/Qsum, which is given to the left of each image. Increasing Qc/Qsum narrows the jet of the inner fluids being sheared through the orifice, creating increasingly small droplets. Please click here to view a larger version of this figure.
Figure 5. Histograms of double emulsion droplets sizes at different flow rates. The average coefficient of variation of the diameter of emulsion droplets produced at a given set of flow conditions is 5.2%. Please click here to view a larger version of this figure.
Figure 6. Droplet diameter versus normalized flow rate parameter. Adjusting the flow rate of the continuous phase allows production of double emulsions that are 30%-100% of the orifice diameter. Please click here to view a larger version of this figure.
Qi [μl/hr] | Qm [μl/hr] | Qsum [μl/hr] | Qc [μl/hr] | Qc / Qsum |
100 | 250 | 350 | 1050 | 3 |
100 | 250 | 350 | 2100 | 6 |
100 | 250 | 350 | 3850 | 11 |
100 | 250 | 350 | 5950 | 17 |
100 | 250 | 350 | 8050 | 23 |
100 | 250 | 350 | 10150 | 29 |
100 | 250 | 350 | 11900 | 34 |
100 | 250 | 350 | 17150 | 49 |
100 | 250 | 350 | 19950 | 57 |
Table 1. Flow rate parameters used for the experiments. The inner phase and middle phase flow rates (Qi, Qm) are held constant, giving a constant combined flow rate (Qsum). The carrier phase flow rate (Qc) is varied to produce double emulsions with different diameters. The ratio Qc/Qsum is the main nondimensional parameter describing experimental conditions.
The double emulsion generating geometry described here is designed to mimic the physics of glass capillary devices8. In these, aligned cylindrical glass capillaries are used to create a three phase coaxial jet that is sheared into uniform double emulsion droplets. The function of our 3D PDMS device is dependent on the central alignment of small features formed with 50 µm tall fabrication with carrier phase channels that are 320 µm in total height. There is a significant potential for to misaligning the taller features patterned by the 2nd layer mask in step 1.7 with respect to the 50 µm tall geometry if masks are not accurately aligned. Proper alignment can be aided by designing alignment marks, such as concentric circles into masks to be co-located during photo patterning. The plasma bonding of the two PDMS halves of the device is a second process that can lead to significant misalignments of the final device. Plasma bonding of PDMS to PDMS is generally instantaneous, so in step 2.4 we describe the wetting of a device surface with DI water to retard bonding and allow manipulation so that the alignment frames shown in Figure 3A can be allowed to lock. If this is attempted without sufficient wetting, the PDMS surfaces will irreversibility bond before coming into proper alignment, and device must be disposed of and new PDMS molds made.
The double emulsion device is designed to take advantage of the fabrication techniques that that lead to uniformly hydrophobic surface properties. However, operation outside of the parameters described in the protocol requires some understanding of the required fluidic processes. A the junction of the inner and middle phases (Figure 1D, point "1"), a relatively high flow of the inner phase and low flow of the middle phase create a two phase jet, with the hydrophobic middle phase coating the channel walls. If the proportionate flow of the middle phase is increased, the generation of discrete water-in-oil droplets will begin to occur, eliminating the ability to form a coherent three-phase jet for drop formation in the orifice (Figure 1D, point "3"). After the channel expansion (Figure 1D, point "2"), a significant amount of carrier phase flow is required to create geometric separation between the middle phase and the hydrophobic channel walls. Reductions in carrier phase flow will eventually lead to the middle phase wetting the hydrophobic device walls. Significant reductions in the carrier phase flow may create flow conditions that are insufficient to shear the inner phases into a long, thin filament, thereby radically altering the physics of double emulsion droplet formation.
Once built, this device is designed to produce double emulsions from 14 to 50 µm, a convenient size for sorting using commercial FACS instruments. If double emulsions outside this size range are desired, the orifice dimensions need to be scaled from the 50 µm x 50 µm size used here. Because the device is designed to produce water / oil / water double emulsions with uniformly hydrophobic surface properties, oil / water / oil double emulsions cannot created unless there were a surface treatment applied to make the device uniformly hydrophilic.
This work demonstrates an easy to fabricate PDMS device capable of the robust formation of water / oil / water double emulsions. Although previous investigators have reported the formation of double emulsions in devices created using 3D lithography14,15, the double emulsions formed in their devices had diameters that were measured in 100s of µm. The device reported here is suited to producing double emulsions an order of magnitude smaller than this, providing volumes similar to mammalian cells and well suited to sorting by FACS.
Although these results may also be attained using glass capillary microfluidics, fabricating glass devices is laborious and requires many hands-on steps per device. For our all PDMS device, fabrication largely consists of molding, bonding, and baking PDMS slabs, processes that are simple, repeatable, and easy to scale to large numbers.
The utility of a lithographically fabricated device to generate double emulsions using co-axial flow focusing has been demonstrated. We hope that the straightforward fabrication and robust functionality of this double emulsion generator design should lead to its adaptation for scientific and industrial applications. Investigators previously deterred by the specialized skills required to work in glass capillary microfluidics, should be more comfortable using PDMS soft lithography, now a common laboratory technique. Furthermore, the small droplet sizes that can be produced is well suited to perform cell and biological assays in droplets, and quantification and sorting using FACS. For industrial applications, it has already been shown that these types of devices can be fabricated into arrays and parallelized10, enabling double emulsion generation rates to increase by orders of magnitudes as compared to single devices. In addition, the ability to form small double emulsions in large coaxial flow focusing channels should make the device resistant to fouling and clogging, which is critical when parallelizing the devices intended to run for long durations without intervention.
The authors have nothing to disclose.
This work was supported by a Research Award from the California Institute for Quantitative Biosciences (QB3), the Bridging the Gap Award from the Rogers Family Foundation, the UCSF/Sandler Foundation Program for Breakthrough Biomedical Research, a grant from BASF, and the NSF through the Faculty Early Career Development (CAREER) Program (DBI-1253293).
Name | Company | Catalog Number | Comments |
Photomasks | CadArt Servcies | ||
3" silicon wafers, P type, virgin test grade | University Wafers | 447 | |
SU-8 3035 | Microchem | Y311074 | |
SU-8 2050 | Microchem | Y111072 | |
Sylgard 184 silicone elastomer kit | Krayden | 4019862 | |
1 ml syringes | BD | 309628 | |
10 ml syringes | BD | 309604 | |
27 gaugue needles | BD | 305109 | |
PE 2 polyethylene tubing | Scientific Commodities, Inc. | B31695-PE/2 | |
Novec 7500 | Fisher Scientific | 98-0212-2928-5 | Commonly knowns as HFE 7500 |
Biocompatable surfactant | Ran Biotechnologies | 008-FluoroSurfactant | |
35,000 MW PEG | Sigma Aldrich | 1546660 | |
Tween 20 | Sigma Aldrich | P1369 | |
Sodium dodecyl sulfate | Sigma Aldrich | L3771 |