To expand the ability of laboratories worldwide to assess the eligibility of patients with lung cancer for treatment with pembrolizumab, in a reliable and reproducible manner, we developed an assay that uses the 22C3 antibody concentrate on a widely available immunohistochemical autostainer, for both biopsy and cytology specimens.
Pembrolizumab monotherapy has been approved for the first- and second-line treatment of patients with PD-L1-expressing advanced non-small cell lung cancer (NSCLC). Testing for PD-L1 expression with the PD-L1 immunohistochemistry (IHC) 22C3 companion diagnostic assay, which gives a tumor proportion score (TPS), has been validated on tumor tissue. We developed an optimized laboratory-developed test (LDT) that uses the 22C3 antibody (Ab) concentrate on a widely available IHC autostainer for biopsy and cytology specimens. The PD-L1 TPS was evaluated with 120 paired whole-tumor tissue sections and biopsy samples and with 70 paired biopsy and cytology samples (bronchial washes, n = 40; pleural effusions, n = 30). The 22C3 Ab concentrate-based LDT showed a high concordance rate between biopsy (~100%) and cytology (~95%) specimens when compared to PD-L1 IHC expression determined using the PD-L1 IHC 22C3 companion assay at both TPS cut points (≥1%, ≥50%). The optimized LDT presented here, using the 22C3 Ab concentrate to determine the PD-L1 expression in both tumor tissue and in cytology specimens, will expand the ability of laboratories worldwide to assess the eligibility of patients with NSCLC for treatment with pembrolizumab monotherapy in a reliable and reproducible manner.
Recent clinical trials have demonstrated the efficacy of pembrolizumab, a humanized monoclonal IgG4 kappa isotype antibody that blocks the interaction between programmed cell death 1 (PD-1) and its ligands, PD-L1 and PD-L2, in the treatment of patients with advanced NSCLC1,2,3,4.
Currently, pembrolizumab is approved for treatment of PD-L1-expressing NSCLC in both treatment-naive patients with a PD-L1 expression TPS of ≥50% and no epidermal growth factor receptor (EGFR) or anaplastic lymphoma kinase (ALK) genomic tumor aberrations3 and for previously treated patients with a PD-L1 TPS of ≥1%1.
PD-L1 protein expression detected by IHC has been widely used as a predictive biomarker assay for anti-PD-1/PD-L1 therapies. In pembrolizumab clinical trials, the PD-L1 TPS obtained with formalin-fixed paraffin-embedded (FFPE) tissue samples was determined using the PD-L1 IHC 22C3 companion assay5. This assay has been approved by the US Food and Drug Administration (FDA) and has been CE-marked in Europe for the determination of the tumor PD-L1 TPS5.
Additional global options across institutions for reliable and high-quality evaluations of the PD-L1 TPS with LDTs which use the 22C3 antibody concentrate are essential to support clinical decisions made regarding patient eligibility for pembrolizumab treatment. A large number of pathology laboratories do not have access to the companion diagnostic PD-L1 IHC 22C3 assay. Therefore, the development of reliable and consistent LDTs compatible with additional, widely available IHC autostainer platforms is essential.
Moreover, there is a need to establish LDTs using cytology samples that are the only specimen type frequently available from NSCLC patients. The PD-L1 IHC 22C3 companion assay is validated for resections, core needle biopsies, and bronchoscopies only if the bronchoscopy yields 100 tumor cells. Although the above sample types are frequently obtained, cytology samples are more easily collected and are the most commonly available sample type in some institutions6,7. However, there is currently no validated diagnostic assay available for the evaluation of the PD-L1 expression in cytology samples; reliable LDTs compatible with cytology samples would further facilitate high-quality PD-L1 testing.
Furthermore, when establishing the clinical validation of an LDT, the IHC should be performed in a similar way to the corresponding clinically validated commercial test8. For instance, several critical steps should be verified to obtain the same signal in serial sections such as antibody titration, pretreatment delays, incubation time, and amplification systems9.
We recently developed an optimized LDT that uses the 22C3 antibody concentrate to evaluate the PD-L1 expression on tumor biopsies and cytology samples10,11. We found a high concordance with the LDT versus the "gold standard" PD-L1 IHC 22C3 assay10,11. This clinically validated protocol will support reliable, high-quality PD-L1 testing across regions globally.
All procedures have been approved by the local ethics committee (Human Research Ethics Committee, Centre Hospitalier Universitaire de Nice/Tumorothèque BB-0033-00025).
NOTE: This protocol is specifically adjusted for the use of the 22C3 antibody concentrate on a commercially available automated IHC stainer (referred to as autostainer here, see the Table of Materials) for tumor biopsies and cytology samples.
1. Preparation of Tumor Tissue Samples
2. Preparation of Cytology Samples
3. PD-L1 Staining Assay
4. Interpretation of the PD-L1 Staining
NOTE: A qualified pathologist should perform the interpretation of the PD-L1 IHC test.
Using the procedure presented here, and as described in detail in this group's recent publications10,11, the optimized LDT was clinically validated with 120 archival FFPE NSCLC biopsy samples from patients who underwent surgical resection or a biopsy at the Pasteur University Hospital, Nice, between March 2007 and March 2016. Moreover, for the evaluation of PD-L1 expression of cytology samples, TPS was evaluated in 70 paired tissue biopsy samples and cell blocks that were prepared from bronchial washes (n = 40) or pleural effusions (n = 30) (collected at the Pasteur University Hospital, Nice, between July 2014 and November 2016). All the slides were freshly cut and stained within 24 hours.
A representative staining pattern with biopsy specimens using the 22C3 antibody concentrate (the LDT), compared with the PD-L1 IHC 22C3 companion assay (gold standard) is shown in Figure 1A and 1B. Using a TPS of ≥1%, 54/120 cases (45%) were PD-L1 positive, while using a TPS of ≥50%, 29/120 cases (24%) were considered positive for PD-L1.
The intraclass correlation coefficient (ICC) used to measure the correlation of the TPS score as a continuous variable was 99% between the LDT and the gold standard. Using both the TPS cut points of ≥1% and ≥50%, the κ scores for interpathologist agreement were equal to 1 within the LDT platform.
A representative staining pattern with cytology specimens using the 22C3 antibody concentrate on the LDT compared with the PD-L1 IHC 22C3 kit (gold standard) is shown in Figure 1C and 1D. The concordance rate of 70 pairs of biopsy and cytology samples with the LDT using either one of the TPS cut points of ≥1% and ≥50% was greater than 95%, and the ICC using the TPS score as a continuous variable was between 0.88 and 0.90. This finding was consistent across each type of cytology sample (pleural effusion vs. bronchial wash) or tumor histology (adenocarcinoma vs. squamous cell carcinoma) with an ICC between 0.82 and 0.96. When comparing the PD-L1 TPS of 37 (out of 70) pairs of cytology samples with the LDT to the biopsy samples with the gold standard, the concordance rate using either the ≥1% TPS or the ≥50% TPS cut point was greater than 97% and the ICC was between 0.93 and 0.95.
Figure 1: Representative staining pattern on biopsy and cytology specimens using the 22C3 antibody concentrate (LDT), compared with the PD-L1 IHC 22C3 kit (gold standard). (A) This panel shows the PD-L1 IHC 22C3 assay used on a tumor biopsy. (B) This panel shows the optimized LDT using the 22C3 antibody concentrate on serial sections from the same tumor biopsy as analyzed in panel A. (C) This panel shows the PD-L1 IHC 22C3 assay in a cell block. (D) This panel shows the optimized LDT using the 22C3 antibody concentrate on serial sections from the same cell block as analyzed in panel C. Please click here to view a larger version of this figure.
We have validated an optimized LDT using the 22C3 PD-L1 antibody concentrate, by comparing it with the corresponding clinically validated commercial test10,11. The 22C3 concentrated antibody-based LDT showed a high concordance rate between biopsy (~100%) and cytology (~95%) specimens when compared to the PD-L1 IHC expression determined using the PD-L1 IHC 22C3 assay at both ≥1% TPS and the ≥50% TPS cut points. As recently recommended by the International Association for the Study of Lung Cancer, the PD-L1 positive areas should be the same for approximately 10 PD-L1 negative samples, 10 PD-L1 positive samples, and 20 samples covering the linear dynamic range of the clinically validated PD-L1 IHC test8. We performed a validation study on 120 biopsies and 70 cytology samples. Overall, the concordance was high, independently of the TPS cutoff for positivity, the type of specimens, or tumor histology. The findings presented here are in agreement with those of other previous studies showing a high concordance rate for both tissue and cytology specimens12,13,14,15.
In this study, all the specimens were fixed in 10% NBF. We did not evaluate the impact of other fixatives, while the effect on PD-L1 staining of other non-formalin fixatives such as alcohol-based fixatives is currently unexplored. The presence of immune cells expressing PD-L1, particularly macrophages, warrants a careful interpretation of the cytology specimens. These samples must be evaluated with reference to the serial hematoxylin and eosin slide, and, in some difficult cases, complementary stains to assess immune cells may be performed to exclude any misinterpretation of the PD-L1 expression in tumor cells only.
This study holds a number of limitations, including a retrospective analysis at a single institution, and the fact that no patient was treated with pembrolizumab to evaluate the clinical outcome. In addition, a minor limitation of the LDT presented here concerns the granular staining pattern that was observed occasionally when using amplification systems, which may make their interpretation more difficult. Moreover, the staining pattern of immune cells is somewhat more intense than that observed with the gold standard. The training of pathologists is necessary to obtain a correct interpretation of PD-L1 staining with IHC.16
In the clinical setting, the robustness of LDTs needs to be maintained over time by seeking certification and accreditation, by following standard operating procedures, and by regularly participating in external quality assessment schemes8. The guarantee of the clinical predictive performance can be ensured only when these prerequisites are accomplished8.
The protocol presented here addresses the critical need for PD-L1 LDTs on both biopsy and cytology samples when analyzed on a widely available autostainer across geographic regions that may not be equipped with the gold standard assay. The LDT presented here, which was optimized using the 22C3 antibody concentrate, can be used to evaluate the PD-L1 expression on both biopsy and cytology samples from NSCLC patients. This will significantly expand the number of laboratories that can offer high-quality PD-L1 testing to identify patients with NSCLC who are eligible for treatment with pembrolizumab monotherapy, in a reliable and reproducible manner.
A potential future direction is to assess the feasibility of using other cytology specimen types (e.g., specimens from fine-needle biopsies, bronchoscopy-guided FNA, endobronchial ultrasound-guided FNA, and bronchial brushes) with 22C3 antibody concentrate-based LDTs.
The authors have nothing to disclose.
This study was sponsored by Merck & Co., Inc., Kenilworth, NJ, USA. The funders played no role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Name | Company | Catalog Number | Comments |
NovaPrep HQ1 | Novacyt | NA | Preservative solution for cytology specimens |
Novaprep® HQ+ B | Novacyt | NA | Mucolytic solution |
Tissue-Tek VIP 6 | Sakura | 6042 VIP 6 | |
Tissue-Tek TEC 5 Tissue Embedding Console System | Sakura | 5229 TEC 5 | |
microscope glass slide SuperFrost Plus | Thermo Fisher Scientific | 4951PLUS4 | |
DL-Dithiothreitol powder | Sigma-Aldrich | D3801 | |
Heidolph Multi Reax Vortex Shaker | Thermo Fisher Scientific | 13-889-410 | |
Shandon Cytoblock | Thermo Fisher Scientific | 7401150 | |
22C3 anti-PD-L1 concentrate antibody | Agilent Dako | #M365329 | |
PD-L1 IHC 22C3 pharmDx | Agilent Dako | SK006 | |
Autostainer Link 48 | Agilent Dako | AS480 | |
BenchMark ULTRA autostainer | Ventana | #750-600 | |
OptiView HQ Universal Linker | Ventana | #760-700 | |
OptiView HRP Multimer | Ventana | #760-700 | |
OptiView Amplification H2O2 | Ventana | #760-099 | |
OptiView Amplifier | Ventana | #760-099 | |
OptiView Amplification Multimer | Ventana | #760-100 | |
OptiView DAB | Ventana | #760-700 | |
OptiView Copper | Ventana | #760-700 | |
Hematoxylin II | Ventana | #790-2208 | |
Bluing Reagent | Ventana | #760-2037 | |
Cell Conditioning 1 (CC1) | Ventana | #950-124 | |
Tissue-Tek Film Coverslipper | Sakura | 4742 |