An Affinity Chromatography Technique for the Purification of Monoclonal Antibodies

-- views • 1:15 min

Transcript

Take an affinity chromatography column containing agarose beads conjugated to Protein A, an antibody-binding protein.

Add a binding buffer with a neutral pH to equilibrate the column.

Load the culture supernatant from an antibody-producing cell culture, containing monoclonal antibodies and contaminating cellular proteins.

The neutral pH facilitates protein A binding to the Fc region of the antibodies while the contaminants flow through.

Measure the UV absorbance of the flow through to create a chromatogram. A high absorbance during sample loading indicates unbound proteins in the flow through.

Wash the column with the binding buffer to remove non-specifically bound contaminants.

Add an elution buffer with a low pH to disrupt the Protein A-antibody interaction. Collect the eluted antibodies by monitoring the peak in absorbance.

Neutralize the pH to prevent antibody degradation and transfer into a centrifugal filter.

Centrifuge to concentrate the antibodies and resuspend in a buffer for downstream assays.

article

01:48

An Affinity Chromatography Technique for the Purification of Monoclonal Antibodies

Related Videos

42 Views

article

11:13

A GPC3-targeting Bispecific Antibody, GPC3-S-Fab, with Potent Cytotoxicity

Related Videos

8.5K Views

article

07:53

Activated Cross-linked Agarose for the Rapid Development of Affinity Chromatography Resins - Antibody Capture as a Case Study

Related Videos

9.2K Views

article

07:32

Bacterial Expression and Purification of Human Matrix Metalloproteinase-3 using Affinity Chromatography

Related Videos

4.4K Views

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved