Method Article
An innovative method for fabricating microfluidic devices using polyethylene terephthalate (PET) lamination significantly reduces the cost and complexity of entrapping and imaging multiple live zebrafish embryos.
Zebrafish embryos are transparent and thus uniquely suited for noninvasive intravital imaging of fundamental processes, such as wound healing and immune cell migration. Microfluidic devices are used for entrapment to support long-term imaging of multicellular organisms, including zebrafish. However, the fabrication of these devices using soft lithography requires specialized facilities and competency in 3D printing, which may not be accessible to every lab. Our adaptation of a previously developed low-cost polyethylene terephthalate lamination method for constructing microfluidic devices increases accessibility by enabling design fabrication and iteration for a fraction of the technical investment of conventional techniques. We use a device made with this method, the Rotational Assistant for Danio Imaging of Subsequent Healing (RADISH), to accommodate drug treatment, manual wounding, and long-term imaging of up to four embryos in the same field of view. With this new design, we successfully capture gross morphological characteristics of the calcium signal around laser ablation and manual transection wounds for multiple embryos in the 2 h immediately following injury, as well as neutrophil recruitment to the wound edge for 24 h.
The ability to respond adequately to injury is critical to the survival of every organism at every scale, from a single cell to multiple tissues. Wounding and its associated responses, such as phagocyte recruitment to damaged areas1 are, therefore, significant topics in cell and tissue biology. Wounds are sensed immediately after the tissue barrier is breached, causing a tissue gradient response involving wound contraction2,3 that coordinates wound healing and subsequent regrowth4. Due to the mechanical nature of this contraction, tools used during experimentation must not physically impede the movement of cells near the injury site.
Zebrafish embryos are an excellent model for studying development and disease, including the vertebrate wound response, due to their ease of care, genetic tractability, and optical transparency5,6. However, immobilizing the whole organism for an extended period is needed to study the longer-term behavior of cells near an injury site. Embedding embryos in low-melt agarose is sufficient for short-term imaging of unwounded tissue. However, the surrounding gel matrix restricts the contraction and relaxation of wounds and impedes the growth of the developing embryos over time7,8,9.
Microfluidics devices can immobilize zebrafish embryos at different developmental stages10,11,12,13,14,15,16, and some, such as the zWEDGI7, accommodate manual wounding. However, there are several drawbacks to available device designs. For example, many devices printed directly on plastic are incompatible with imaging on inverted microscope systems due to poor optical transparency. Furthermore, parallel channels allow multipositional imaging10,11, but the physical movement of the microscope stage introduces lag in image acquisition. Repeated iteration and optimization to solve these issues are difficult when considering the financial and technical investment required for every new design, especially for current gold-standard methods of device fabrication using polydimethylsiloxane (PDMS) soft lithography. The very first step, the creation of a master mold, often requires knowledge of 3D modeling software, access to specialized fabrication equipment, and (depending on the material used) an additional antiadhesion treatment such as silanization before the mold is ever used. Curing the PDMS once poured takes at least one hour at high temperatures, and generally, must be done under a vacuum or with a clamp for best results, often in a clean room7,17,18. As in developmental biology, these costs can rapidly prove prohibitive for experiments requiring many unique microfluidic environments across multiple models.
Of available alternative construction materials, polyethylene terephthalate (PET) is durable, nontoxic, and easy to manipulate. PET sheets are widely available plastic lamination pouches carried at most office supply stores, with pre-applied thermally-activated adhesive (generally ethylene vinyl acetate). By shaping these PET sheets using commercially available craft cutters (i.e., computer-controlled cutting plotters designed for home crafters) and adhering the layers to one another using standard thermal lamination equipment, a wide range of potential designs can be quickly generated and iterated. We have therefore adapted a previously described method of design and construction involving stacked PET18 to create the Rotational Assistant for Danio Imaging of Subsequent Healing (RADISH) (Figure 1A,B). A rotational arrangement of several approximately wedge-shaped restraining channels optimizes proximity while allowing room for manual tail fin transection, accommodating imaging immediately after wounding and simultaneous long-term imaging of multiple wounded zebrafish embryos within the same field of view. Additionally, this construciton method drastically reduces the upfront capital costs and the time required for device construction while maintaining reusability.
This study used 3 days post fertilization (dpf) embryos but can be designed to use 2-14 dpf embryos. The zebrafish experiment was conducted by internationally accepted standards. The Animal Care and Use Protocol was approved by The Purdue Animal Care and Use Committee (PACUC), adhering to the Guidelines for using Zebrafish in the NIH Intramural Research Program (protocol number: 1401001018).
1. Assembly of the PET microfluidics device
NOTE: Any part of this protocol may be paused at any time except for steps 1.8 and 1.9, which are sensitive to the drying time of cyanoacrylate.
2. Positioning of the embryos within the RADISH and preparation for imaging
3. Imaging of calcium transients after wounding
NOTE: This section and Section 4 are optional and provided as potential sample experiments.
4. Imaging of neutrophil recruitment after wounding
For comparison of image quality produced by a laser scanning confocal with and without the aid of the RADISH, embryos expressing the intensiometric calcium biosensor GCaMP6f22,24 in the outer epithelial layer were stained using MemGlow 560 and imaged with an inverted laser scanning confocal covering the entire thickness of the tail fin fold at a time interval of 1 min per frame (Video 2). Without the device (Figure 3A), samples drifted significantly and displayed banding artifacts, resulting from limiting the Z-stack to 25 slices to fit within the specified temporal resolution compared to imaging with the device (Figure 3B). The RADISH was further used to examine calcium transients immediately after wounding in embryos expressing Tg(krt4:Gal4, UAS:GCaMP6f) x (cdh1-dTomato)xt1821, highlighting endogenous E-cadherin in red and cytoplasmic calcium in the apical epithelium in green. Embryos were wounded parallel to the sides of the central chamber using a scalpel blade (Figure 4A and Video 3) or by direct laser ablation with an IR laser (Figure 4B), then imaged through the full thickness of the tail fin fold.
To demonstrate the effectiveness of the RADISH for longer-term imaging, we investigated immune cell recruitment to the wound edge for 24 h after wounding. Four zebrafish embryos at 3 dpf expressing Tol2(mpx:Dendra2) to label neutrophils in green23 were wounded by tail transection and imaged for 24 consecutive hours (Figure 5). Imaging artifacts typically associated with late 3 dpf to 4 dpf embryos, such as tilting and floating, were notably absent (Video 4), and tissue regrowth continued unimpeded for the experiment. Neutrophil recruitment to the wound edge continues throughout the image (Figure 5B), with most migrating cells leaving by 24 h post wounding (Figure 5C), consistent with expected inflammation resolution25,26.
Figure 1: Design and use of PET laminate microfluidics device (RADISH). (A) Schematic illustrating individual PET layers' arrangement and thickness, assembly, and device mounting and use. (B) Final product, mounted in a ¾" cutout in a 35 mm dish. (C) 3 dpf embryos, (D) 5 dpf embryos, (E) 14 dpf embryos housed inside the device. Scale bars = 1,000 µm (C,D,E). Abbreviations: PET = polyethylene terephthalate; RADISH = Rotational Assistant for Danio Imaging of Subsequent Healing; dpf = days post fertilization. Please click here to view a larger version of this figure.
Figure 2: Step-by-step photographic illustration of PET device assembly. (A) Individual PET layers before lamination. (B) PET layer secured to 25 mm coverslip using office tape. (C) Device-in-progress secured to a plastic backing for lamination. (D) Completed PET laminate device. (E) Modified 35 mm dish before device mounting. (F) Completed device. Abbreviation: PET = polyethylene terephthalate. Please click here to view a larger version of this figure.
Figure 3: Improved quality of laser scanning confocal images of 3 dpf embryos in the RADISH design. (A) Representative image of 3 dpf Tg(krt4:Gal4, UAS:GCaMP6f) embryo stained with MemGlow 560 without using the RADISH. (B) Representative image of Tg(krt4:Gal4, UAS:GCaMP6f) embryo stained with MemGlow 560 mounted in the RADISH. Scale bar = 100 µm. Abbreviations: RADISH = Rotational Assistant for Danio Imaging of Subsequent Healing; dpf = days post fertilization. Please click here to view a larger version of this figure.
Figure 4: Compatibility of the RADISH design with multiple wounding techniques. (A) 3 dpf Tg(krt4:Gal4, UAS:GCaMP6f) x (cdh1-dTomato)xt18 embryo wounded via manual tailfin transection. (B) Embryo wounded via laser ablation. (C,D) The same embryos 4 min later, showing the progression of calcium transients after wounding. Scale bars = 100 µm. Abbreviations: RADISH = Rotational Assistant for Danio Imaging of Subsequent Healing; dpf = days post fertilization. Please click here to view a larger version of this figure.
Figure 5: Long-term entrapment of embryos for imaging of phagocyte migration. (A) Tol2(mpx:Dendra2) embryos immediately post wounding, and the same embryos (B) 25 min, (C) 12 h, and (D) 24 h later. Scale bars = 100 µm. Please click here to view a larger version of this figure.
Supplemental File 1: Vector and PNG files for RADISH device design as shown in Figure 1A. Please click here to download this File.
Video 1: Brightfield video of manual tailfin transection using RADISH device design. Abbreviation: RADISH = Rotational Assistant for Danio Imaging of Subsequent Healing. Please click here to download this Video.
Video 2: Comparison of unmounted versus mounted 3 dpf Tg(krt4:Gal4, UAS:GCaMP6f) embryos from Figure 3, acquired at 1 frame/min. Please click here to download this Video.
Video 3: Calcium transients in Tg(krt4:Gal4, UAS:GCaMP6f) x (cdh1-dTomato)xt18 embryos wounded via scalpel transection or laser ablation from Figure 4, acquired at 1 frame/min. Please click here to download this Video.
Video 4: Long-term imaging of leukocyte recruitment and tail fin regeneration in Tol2(mpx:Dendra2) embryos. Abbreviations: RADISH = Rotational Assistant for Danio Imaging of Subsequent Healing; hpw = hours post wounding. Please click here to download this Video.
The core premise of the PET lamination method involves cost reduction and minimization of technical barriers to entry compared to traditional means of creating microfluidic devices, such as soft lithography or PDMS molding. As such, the only critical steps in this protocol are the accurate alignment of PET layers during lamination and waterproofing of the device edges after construction. All other parts of the method may be modified according to experimental needs (e.g., if immobilization of younger or older embryos is required, the thickness of the PET layers can be adjusted to narrow or widen the channel accordingly). Designs can be created in any shape and laminated onto any base material as long as the material is heat-safe and thin enough to be sent through the laminator18,27. As PET sheets are cut using commercially available craft cutters, which read two-dimensional rather than three-dimensional image files, the design and iteration of any given device require only the ability to manipulate black and white shapes in an image editing program, dramatically decreasing the difficulty. This flexibility in software choice and the commercial availability of materials used represent a dramatic increase in the accessibility of microfluidics devices in general, enabling their use in any lab or classroom for an upfront cost of less than $500 at the time of writing this paper, compared to over $2000 in starting equipment costs for PDMS mold fabrication17.
Furthermore, the time required to create a single useable device using this method is dramatically reduced compared to typical soft lithography. Assuming familiarity with the requisite software, one unique PET laminate microfluidic device can be conceptualized, manufactured, and immediately used in ~3 h, including the curing time of cyanoacrylate. Iteration takes even less, depending on the scale of edits needed to the original design files, and multiple variant devices can be made in parallel with minimal additional time invested and no extra equipment. Although devices created using this method are less durable than those made with PDMS for a variety of reasons (including adhesive failure and internal stresses for shapes with many layers and long aspect ratios), their inexpensive nature and scalable fabrication mean that they are far more easily replaced. As commercial craft cutters are marketed with the ability to repeat the same precise cut over and over and PET has no expiration date, the individual layers of a completed design can be pre-prepared in bulk for assembly and use at any point in the future, further reducing the overall time investment required.
However, speed and variability come at the cost of precision. The main limitation of PET lamination compared with traditional PDMS molding is spatial resolution. Commercial craft cutters often struggle with sharp angles on shapes smaller than approximately 500 µm, and PET lamination sheets are not available at thicknesses under 33 µm (not including the thickness of the adhesive itself, which is not included in the measurement on the package), significantly limiting the scale at which this technique can be applied. The maximum number of layers is also restricted by the size of the laminator, which may vary from manufacturer to manufacturer.
The final version of the protocol, including variations in registration and alignment technique, materials suppliers, and software, will largely depend on user preference. However, the method's accessibility and overall flexibility reduce the cost of adaptation. Individual users are encouraged to iterate as much as necessary to adapt the approach to their unique needs.
The authors have no conflicts of interest to disclose.
The work was supported by research funding from the National Institutes of Health (R35GM119787 to QD). This work is based upon efforts supported by EMBRIO Institute, NSF contract #2120200, a National Science Foundation (NSF) Biology Integration Institute. Confocal imaging was performed at the Purdue Imaging Facility. We thank Dr. Guangjun Zhang (Purdue University) for providing the Tg(UAS:GCaMP6f) line. We thank Dr. David Tobin (Duke University) for providing the (cdh1-tdtomato)xt18 line.
Name | Company | Catalog Number | Comments |
25 mm diameter round coverglass #1 thickness | Chemglass Life Sciences | CLS-1760-025 | Base mounting material. Thickness was chosen based on the specifications of the microscope lens. |
Adobe Illustrator v28.5 | Adobe | N/A | Vector editor for designing the device pattern. |
Calcium Chloride Dihydrate | Fisher | C79 | For making E3 medium. |
Design PNG Files (Online Mirror) | N/A | N/A | https://i.ibb.co/QPYj7BL/multipositionalv7-1.png https://i.ibb.co/FDpQYXc/multipositionalv7-2.png https://i.ibb.co/5TG4SB6/multipositionalv7-3.png |
Design Space for Desktop v8.39 | Cricut | N/A | Proprietary software to drive the craft cutter. |
Fusion Plus 7000L | GBC | 1703098 | Thermal laminator. This specific model was selected for quality of life features, such as maximum sheet stack size. |
German Carbide Premium Blade | Cricut | 10396595 | Replacement blade for craft cutter. |
Gray Basic Tool Set | Cricut | 10307854 | Tool set for weeding cuts, cleaning mats, and mounting PET sheets. Optional. |
Magnesium Chloride Hexahydrate | Acros Organic | 223211000 | For making E3 medium. |
Maker 3 | Cricut | 10669040 | Craft cutter. The Cricut Maker 3 was selected over the Silhouette Cameo 4 for software user friendliness, but any craft cutter capable of reading black and white images will work. |
MemGlow 560 | Cytoskeleton | MG02-10 | Red live cell dye for cell membranes. No washing needed. Used in Figure 3. |
No. 22 Carbon Scalpel Blade | Surgical Design | 22-079-697 | Scalpel blade for manual cut adjustments. Any similar craft knife (e.g., X-acto #2 knife) will also work. |
Potassium Chloride | Fisher | BP366 | For making E3 medium. |
Sodium Chloride | Fisher | BP358 | For making E3 medium. |
StandardGrip Adhesive Cutting Mat | Cricut | 10138842 | Mounting mat for craft cutter. The LightGrip cutting mat will also work, but avoid the StrongGrip and FabricGrip cutting mats as the strength of the adhesive may warp the final cut during weeding and removal. Adhesive on mats will eventually wear out with continued use; replace mats as needed. |
Stationery Tape 12 mm | Deli | 30014 | Office tape. Deli brand was chosen via testing for ease of removal after lamination. |
Super glue, liquid | Loctite | 1364076 | Cyanoacrylate glue. Glue used must be liquid. Gel formulations will not sufficiently seal or waterproof device edges. |
PET Thermal Laminating Pouches, 3 mil | Scotch | TP3854 | PET sheets with thermal adhesive. Sheets arrive as sandwiched pairs and should be separated before use. Cut with adhesive side facing down. |
PET Thermal Laminating Pouches, 5 mil | Scotch | TP5854 | PET sheets with thermal adhesive. Sheets arrive as sandwiched pairs and should be separated before use. Cut with adhesive side facing down. |
Tricaine (MS-222) | Syndel | IC10310680 | Fish anaesthetic. |