The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence of an analyte but is usually too low to be reliably quantified. For quantification, we need another value called the limit of quantification, which is defined as the lowest quantity of analyte that the instrument can quantify. Its value corresponds to the concentration at which the signal is ten times larger than the standard deviation of the blank signal.
来自章节 1:
Now Playing
Chemical Applications of Statistical Analyses
4.6K Views
Chemical Applications of Statistical Analyses
1.2K Views
Chemical Applications of Statistical Analyses
2.9K Views
Chemical Applications of Statistical Analyses
4.8K Views
Chemical Applications of Statistical Analyses
1.2K Views
Chemical Applications of Statistical Analyses
1.2K Views
Chemical Applications of Statistical Analyses
675 Views
Chemical Applications of Statistical Analyses
4.4K Views
Chemical Applications of Statistical Analyses
302 Views
Chemical Applications of Statistical Analyses
375 Views
Chemical Applications of Statistical Analyses
487 Views
Chemical Applications of Statistical Analyses
333 Views
Chemical Applications of Statistical Analyses
2.9K Views
Chemical Applications of Statistical Analyses
3.3K Views
Chemical Applications of Statistical Analyses
1.5K Views
See More
版权所属 © 2025 MyJoVE 公司版权所有,本公司不涉及任何医疗业务和医疗服务。