Atomization, converting samples into gas-phase atoms and ions, is essential for atomic spectroscopy. The flame temperature required for atomization affects the efficiency of the atomic spectroscopic methods by increasing the atomization efficiency and the relative population of the excited and ground states.
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from 2500 K to 2600 K can increase the population of excited-state sodium atoms by 45%, while the ground-state population decrease is negligible. Since atomic emission spectroscopy (AES) relies on photon emission from these excited states, it is highly temperature-dependent. In contrast, atomic absorption spectroscopy (AAS) and atomic fluorescence spectroscopy (AFS) primarily depend on the ground-state population and have less significant temperature dependence. However, for easily ionizable elements, an increase in flame temperature causes a loss of atoms by ionization, adversely affecting absorption and fluorescence spectral intensity.
In addition, for atomic spectroscopy overall, higher temperature increases the velocity of the atoms, making the Doppler effect more pronounced. This results in the broadening of atomic spectral lines and decreasing peak height.
From Chapter 14:
Now Playing
Atomic Spectroscopy
244 Views
Atomic Spectroscopy
660 Views
Atomic Spectroscopy
500 Views
Atomic Spectroscopy
429 Views
Atomic Spectroscopy
270 Views
Atomic Spectroscopy
307 Views
Atomic Spectroscopy
522 Views
Atomic Spectroscopy
258 Views
Atomic Spectroscopy
551 Views
Atomic Spectroscopy
271 Views
Atomic Spectroscopy
126 Views
Atomic Spectroscopy
426 Views
Atomic Spectroscopy
153 Views
Atomic Spectroscopy
126 Views
Atomic Spectroscopy
202 Views
See More
Copyright © 2025 MyJoVE Corporation. All rights reserved